If the mv signal drops rapidly after the reference air is disconnected, this probe is no longer usable and needs to be returned to the manufacturer for further testing and evaluation.
If it is necessary to remove the oxygen probe from the hot furnace chamber, this operation must be done with extreme caution. In any case, the speed of withdrawal or insertion should not exceed 2 inches per minute.