1. Product Overview
Model: FC8600 (part of Marathon Sensors' Quicksilver series)
Type: High-Temperature Oxygen Sensor (ZrO2), designed specifically for measuring oxygen content in molten metals (such as aluminum, copper, zinc alloys).
Application Scenarios: Casting, metal heat treatment, metallurgical industry, used for real-time monitoring of oxygen partial pressure in furnaces or molten metals to optimize process control.
2. Key Features
Quick Response: The Quicksilver series is renowned for its rapid and stable readings, ideal for dynamic melting environments.
Alloy Compatibility: The probe material is resistant to high-temperature corrosion and suitable for non-ferrous alloys such as aluminum and copper.
Temperature Range: Typically supports 600°C to 1200°C (specific details may vary by model, please refer to manual).
Output Signal: Analog signal (such as 4-20mA or mV) or digital interface, compatible with most controllers.
3. Installation and Maintenance
Installation Points:
Insert to the specified depth in the furnace or metal liquid (refer to the user manual).
Ensure airtightness to prevent air leakage affecting the reading.
Calibration: Regularly calibrated with standard gas (such as N/O2 mixed gas with known oxygen concentration).
Service Life: Depends on the operating environment; continuous operation at high temperatures may shorten the service life.
4. Troubleshooting Common Issues
Unstable readings:
Check if the probe is aging or contaminated.
Ensure the reference air channel is clear (external reference gas required for some models).
No signal output
Verify power/source wiring usually requires 24 DC power supply).
Check if the probe has reached the working temperature (some require external heating preheating).
5. Alternative and Compatibility
Replacement Model: Marathon may offer an updated model (such as FC8700).
Warranty: 12 months