I. Overview
The DWK-500E explosion-proof constant current locatable low flow atmospheric sampler, independently researched and manufactured by Yancheng Dewang Instrument Equipment Co., Ltd., is an instrument with patented technology capable of collecting gaseous or vapor samples from the atmosphere. It is suitable for individual toxic substance sampling of occupational hazards at workplaces and for sampling dust-like substances such as metals; it is also applicable to routine and emergency monitoring of environmental atmosphere, public health, and occupational health.
The Constant Current Locatable Low-Flow Air Sampler is primarily composed of: the operating system, intelligent control system, air pump, flow measurement system, and positioning data upload module. It can operate continuously for over 20 hours on a single charge, convenient for sampling tasks. The instrument features real-time flow, sampling volume, sampling time, temperature, atmospheric pressure, and battery level displays. It boasts a compact structure, small size, lightweight, durability, strong load-bearing capacity, stable and accurate flow rate, and low noise. It is portable for convenient use during sampling.
Section II: Application Fields
The DWK-500E Constant-Current Positionable Low-Flow Sampler is an ideal instrument for short-term or long-term air sampling by environmental protection personnel in disease prevention centers, environmental monitoring stations, colleges and universities, research institutions, and testing organizations.
Standard of Execution:
JJG956-2013规程 for Atmospheric Sampler Calibration
GBZ/T300-2017 Specification for Sampling of Harmful Substances in Workplace Air
GB50325-2020 Standard for Indoor Environmental Pollution Control of Civil Engineering Construction Projects
HJ 194-2017 "Technical Specification for Manual Monitoring of Ambient Air Quality"
Section 3: Technical Specifications
Flow Range: (0-500ml/min) (Flow adjustable)
2. Constant Current Mode: Electronic Digital Constant CurrentSupports various functions including data entry, mobile data collection, historical data download, data logic verification and audit, statistical analysis, and visual presentation.(Equipped with a digital DTU positioning module, it allows for real-time upload of sampling data, fault alarms, and display of sampling location's latitude and longitude.Flow is unaffected by voltage fluctuations and air resistance changes.◆ Anti-siphon liquid particle device: A patent (Patent No.: ZL2020 2 1017091.7) held by our company for a gas sampling equipment, specifically designed to prevent the back-siphonage of absorbent liquid into the instrument during sampling. This device completely blocks the liquid from entering the instrument's interior in case of back-siphonage and features an alert function. When the absorbent liquid reaches the sampling device, a white circular sticker on the waterproofing membrane turns red upon contact with liquid, alerting the user. The instrument will sound an alarm, with the fault light flashing, and automatically shut down after 30 seconds. To resume operation, one only needs to pause the main unit, open the waterproofing device, clean it, replace the new waterproofing membrane, and then continue working. This significantly reduces the instrument's failure rate and effectively streamlines the sampling work process, eliminating the need for factory repairs.
3. Flow Error: ≤±2.5%
4. Traffic Stability: ≤3%
5. Traffic Repetition: ≤2%
6. Timing Error: ≤±0.01%
7. Timing Range: 1min - 99h 59min
8. Operating Temperature: -10~50℃
9. Relative Humidity: ≤95% RH
10. Battery Life: Not less than 20 hours
11. Load Capacity: ≥ 35,000 Pa
12. Sampling Mode: ①Cyclic Sampling (99 cycles), available for cyclic volume or time sampling. Set the sampling time, interval time, and number of cycles.
②Manual sampling, after setting the sampling flow, sampling can be directly initiated, and the timer starts. Manual shutdown is available.
③ Schedule sampling with set sampling times; sampling stops once the time is reached.
④ Constant Volume Sampling (1000L), set to sample at a specific volume.
12. (1) Lock Function: The panel lock feature prevents changes to the settings during the pump's operation.
12 (2) Flow Fault Correction: Fault alarm function display activated when flow variation exceeds ±5% within the back pressure specification range.
If the error cannot be corrected within 30 seconds, the pump will cease operation.
12 (3) The extra-large LCD screen displays: flow rate, sampling time, sampling volume, back pressure, battery estimated sampling time, date, time, operating mode, and battery remaining power.
12(4) Humidity Range: Operating humidity: 5-95%, non-condensing; Storage humidity: 5-98%, non-condensing
12(5) You can view the real-time working data, actual conditions, and latitude and longitude information of the sampler on our provided cloud platform.
13. Main Unit Dimensions: 130*44*95mm (Length*Width*Height)
14. Operating Noise: <55 dB
15. Instrument Weight: 0.52 kg
16. Power: DC 7.4V, 220VAC ±10%
17. Volume Calculation: Real-time Display
18. Data Storage & Download: 1000 sets of real-time queryable sample data, with USB serial port download.
19. Explosion-proof Certificate No.: CNV20.0233 Explosion-proof Mark: Ex ib IIB T4 Gb