I. Product Overview
The CCZ20 Dust Sampler is an instrument used to measure the concentration of suspended dust in the ambient air.
In accordance with GB/T 20964-2007 "General Technical Conditions for Dust Samplers" and JJG520-2005 "Technical Regulations for Dust Samplers"
"Regulation for Instrument Calibration," Q/320582YLA001-2010 Dust Sampler Enterprise Standard, and GB3836.1-2000.
ExibI classification (intrinsically safe) explosion-proof design as per GB3836.4-2000 standard, suitable for underground coal mines.
Workplaces containing other explosive gas hazards.
The sampler is equipped with two dust precollectors, one being a full-process precollector for measuring the total dust concentration.
Another type is a preliminary collector for measuring the concentration of respirable dust, which can effectively trap respirable dust harmful to the human body.
Separate dust and non-respirable dust; one collection can obtain both respirable and non-respirable dust samples.
This, with its separation efficiency reaching the internationally recognized "BMRC" curve standard, is a more reliable dust separator.
The instrument is equipped with a high-performance suction pump featuring pulseless airflow, with great load and negative pressure capabilities, and automatic timing.
Sampling, stable flow, easy operation, and portability.
II. Model and Main Parameters
Rated sampling flow rate: 20 L/min
Sample Flow Rate Range: 10 ~ 25 L/min
Sampling Flow Error: ≤5% FS
Sample Flow Stability: Sample flow stability ≤ 3.0%FS within 30 minutes
Load Capacity: ≥200Pa
Continuous Operation Time: ≥100min
Work Noise: ≤75 dB(A)
Flow meter accuracy: Not less than 2.5 grade.
Sampling Time Error: ≤0.1%
Average flow rate of sampling port: 0.5 to 4.0 m/s
Sampling Accuracy: ≤10%
Safety-Approved Battery Pack: 2.3A/h Ni-H Qing batteries, 6 units, in series with 4 pieces of 0.51Ω/10W resistors
The limiting resistor, encapsulated with epoxy resin, is placed inside the battery case. Rated voltage Ue: 7.2V; its maximum
Output Voltage U0: 8.4V; Maximum Output Current Io: 3.0A.
Dimensions: 149×202×80 (mm)
Weight ≤ 1.75 Kg
Operating Environment: Temp: 0-40℃; Humidity: ≤95%
Atmospheric pressure: 80KPa to 110KPa; contains gas or coal dust explosion hazards in underground coal mines