Basic Function of the Instrument
A workstation for microbial inoculation, isolation, culture, and medium preparation, observation, etc., which can be operated directly within the anaerobic environment of the workstation, without the need to remove the workstation from the anaerobic environment for operation.
Technical Parameters and Specifications
Workstation Material:
1.1 The workstations are made of polypropylene resin boards, featuring stable structure, excellent thermal insulation, and moisture retention properties.
1.2 Panel Connection: The workstations' interior walls are one-piece molded, eliminating the risk of microbial growth in crevices.
2. Workstation Dimensions:
2.1 External Occupied Dimensions: Height ≤ 750mm; Width ≤ 1620mm; Depth ≤ 720mm;
2.2 Internal Dimensions: Height ≥ 480mm; Width ≥ 1000mm; Depth ≥ 500mm;
3. Transfer Case
3.1 Standard transfer box located externally to the studio;
3.2 Transfer Box Dimensions: Height ≥ 300mm; Width ≥ 320mm; Depth ≥ 300mm;
4. Workstation Performance Metrics Control Range:
4.1 Temperature control range: Room temperature +7℃ to 45℃, accuracy 0.1℃.
4.2 Humidity Control Range: Indoor humidity -75%
4.3 Anaerobic Control: Below 0.1%, optional high-efficiency accessories available for achieving an anaerobic level of 5 PPM.
5. Sleeveless Glove System: Operate all delicate tasks at the work station with just standard laboratory gloves, without the need to work through thick gloves.
6. Standard with a removable front panel, the front panel can be easily detached by hand without any tools, facilitating the handling of experimental equipment.
7. The operation room is equipped with fluorescent lights, whose brightness can be adjusted via a touch screen.
8. Equipped with an activated carbon filtration system for removing volatile gases and fatty acids, as well as growth inhibitors.
9. Equipped with a palladium catalytic system, capable of removing trace oxygen from within the work station.
10. Optional 15cm single dish fits into the system, non-drawer style, without push rods, allowing for quick transfer of individual dishes or small vessels.
11. Optional cable sealing ports for connecting power or data cables between workstations and their external sources.
12. Optional waste disposal ports are available for timely waste management during experiments, ensuring the cleanliness of the work station.
13. Temperature, humidity, and low-pressure alarm system