High-Temperature Test Chamber Product Application Scope: Industries
Guangdong Hongjun Instrument Equipment Co., Ltd.'s HJ/CQ series testing equipment is used for testing the heat, cold resistance, high-temperature resistance, and moisture resistance of various products and materials. Suitable for factories producing electronics, silicon chips, batteries, photovoltaic modules, electrical appliances, food, vehicles, metals, chemicals, laptops, LEDs, 5G, mobile phones, communication, aerospace, aviation, semiconductors, chips, diodes, transistors, auto parts, new energy, lithium batteries, building materials, etc., as well as for production workshops, laboratories, and research institutions to conduct temperature and humidity environmental tests on finished and semi-finished products.
II. High-Temperature Test Chamber Specifications:
1. Internal dimensions: 600mm (width) × 600mm (depth) × 600mm (height); (Custom made for non-standard specifications!)
2. External dimensions approximately: 900 (width) × 830 (depth) × 1520 (height) mm; actual item measurement applies.
3. Case Structure: Integrated one-piece design, vacuum pump located at the bottom of the machine.
4. Divider: Internally divided into two layers, consisting of two stainless steel mesh plates, which are removable and can be slid out.
5. Panel Installation: The control panel is installed beneath the machine.
6. Door Opening: Single door opening from right to left, equipped with a viewable window of W300*H350mm.
7. Inner material: 5.0mm thick stainless steel plate; reinforced inner box, vacuum sealed without deformation.
8. Outer material: Cold-rolled steel plate, thickness 1.2mm, powder coated (beige white).
9. Insulation material: high-temperature-resistant rock wool, with excellent thermal insulation performance.
10. Equipped with brake-mounted casters for easy movement.
Section 3: High-Temperature Test Chamber Temperature Control System
1. The temperature controller is a digital display gauge, featuring single-point temperature control, automatic calculation, simultaneous display of PV/SV, and key setting.
2. Timer Function: Equipped with an Omron timer, heating and vacuum extraction will stop upon completion of the set time, along with an alarm prompt.
3. Temperature-sensing wire, Type K
Output: 3-32V
5. The current controller is a PID + SSR solid-state relay.
6. Heating Method: Uses a three-sided radiant heat method (left, right, and bottom sides).
7. Heating materials include: high-temperature stainless steel heating tubes.
8. Temperature Range: room temperature +30℃ to 250℃ (adjustable and controllable in between)
9. Control Accuracy: ±0.5℃
10. Display Accuracy: 0.1℃
11. Internal Temperature Deviation: ±5.0℃ (No vacuum, unloaded test)
4. High-Temperature Test Chamber Vacuum System:
1. Door Sealing Method: The machine door is sealed with a high-density silicone seal.
2. Vacuum Display Method: Vacuum gauge is used to indicate vacuum level.
3. The vacuum control method is: automatic control. The vacuum pump stops working when the vacuum reaches the set value, automatically replenishes vacuum when it falls below the upper limit, and continues this cycle to maintain a constant vacuum state.
4. Pressure Relief Method: After the test, manually activate the pressure relief function for pressure release; controlled by an electromagnetic valve.
5. Vacuum degree range of machine: ≤1.0×10^5 to 1.0×10^1 (Pa), [i.e., 10 Pa]
6. Vacuum Pressure Leak Rate: 0.01 Pa/L per second.
7. Vacuum Tube: Made with stainless steel vacuum corrugated tube.
8. Vacuum Sensor: Utilizes KF16 vacuum resistance gauge tube.
9. Vacuum Setting Method: The vacuum level can be freely set according to requirements, such as for a vacuum level of (100Pa), the gauge is set to (1.0 E 2), for a vacuum level of (10Pa), the gauge is set to (1.0 E 1), and so on. It employs upper and lower limits to control vacuum extraction and stoppage.
10. Vacuum Control Valve: Utilizing KF40 fitting high vacuum with pressure differential valve.
11. Vacuum Pump Configuration: Double-stage direct-connected oil pump used.
12. Vacuum Level Display: Utilizes a resistance vacuum gauge to show the vacuum level, with the display in scientific notation format.
V. Nitrogen Filling Device for High-Temperature Test Chambers
1. Pressure Regulator: Controls nitrogen pressure, typically set to 2-3 kg under normal operation.
2. Flowmeter: Controls the size of nitrogen entering the box, range is 10L/min-60L/min.
3. Nitrogen Filled Pipeline: Nitrogen pipeline enters just behind the box, equipped with a 2-meter long 8mm air tube.
4. Nitrogen inflation port: One nitrogen port inflates into the box.
5. Nitrogen Source: The user company provides the nitrogen source.
Six, High-Temperature Test Chamber Power Supply:
Voltage: AC 380V/50Hz, three-phase five-wire system
2. Current: 14A
3. Power: 9.5 kW
Section 7: High-Temperature Test Chamber Protection System
Over-temperature Protection System: Automatically stops heating and power supply when the temperature exceeds the set over-temperature protection limit, thereby ensuring the safety of both the product and the machine.
2. Circuit Protection Systems: Grounding protection, express insurance, phase loss protection, overload protection, circuit breakers, etc.