One, High Temperature Test Chamber Product Application Scope Industry:
Guangdong Hongjun Instrument Equipment Co., Ltd.'s HJ/CQ series testing equipment is used for testing the heat, cold resistance, high-temperature resistance, and moisture resistance of various products and materials. Suitable for factories producing electronics, silicon chips, batteries, photovoltaic components, electrical appliances, food, vehicles, metals, chemicals, laptops, LEDs, 5G, mobile phones, communication, aerospace, aviation, semiconductors, chips, diodes, transistors, auto parts, new energy, lithium batteries, construction materials, etc., as well as for production workshops, laboratories, and research institutions to conduct temperature and humidity environmental tests on finished and semi-finished products.
II. High-Temperature Test Chamber Specifications:
1. Internal dimensions: 600mm (width) × 600mm (depth) × 600mm (height); (Custom made for non-standard specifications!)
2. External dimensions approx.: 900 (width) × 830 (depth) × 1520 (height) mm; actual product measurement applies.
3. Enclosure Structure: Integrated one-piece design; vacuum pump located at the bottom of the machine platform.
4. Interlayer: Internally divided into two layers, with two stainless steel mesh plates, which are removable and adjustable.
5. Panel Installation: The control panel is installed below the machine.
6. Door Opening: Opens from right to left, equipped with a可视窗 W300*H350mm window on the door.
7. Inner material: stainless steel plate, thickness 5.0mm; reinforced inner box, vacuum-sealed without deformation.
8. Outer material: Cold-rolled steel plate, thickness 1.2mm, powder coated finish (米白色) - Milky white.
9. Insulation material: high-temperature-resistant rock wool, excellent thermal insulation performance.
10. Equipped with brake casters for easy movement.
Section 3: High-Temperature Test Chamber Temperature Control System
1. The temperature controller is a digital display gauge, featuring single-point temperature control, automatic calculation, simultaneous display of PV/SV, and button settings.
2. Timer Function: Equipped with an OMRON timer, heating and vacuum extraction will stop upon completion, along with an alarm notification.
3. Thermocouple Type K;
4. Output voltage: 3-32V
5. The current controller is a PID + SSR solid-state relay.
6. Heating Method: Employing a three-sided radiant heat method; (left, right, and bottom sides)
7. Heating materials include: high-temperature stainless steel heating tubes.
8. Temperature Range: Room temperature to +250°C (adjustable and controllable within this range).
9. Control Accuracy: ±0.5℃
10. Display Accuracy: 0.1℃
11. Inside box temperature deviation: ±5.0℃ (no vacuum, unloaded test)
Section 4: High-Temperature Vacuum Test Chamber System
1. Door sealing method: The machine door uses high-density silicone seal strips for sealing.
2. Vacuum Display Method: Vacuum gauge for vacuum level indication.
3. The vacuum control method is: automatic control. The vacuum pump stops when the vacuum reaches the set value, automatically replenishes vacuum when it falls below the upper limit, and cycles this process to maintain a constant vacuum state.
4. Pressure relief method: After the test, manually initiate the pressure relief function for pressure release; controlled by an electromagnetic valve.
5. Machine vacuum range: ≤1.0×10^5 to 1.0×10^1 (Pa), [i.e., 10 Pa]
6. Vacuum Pressure Leak Rate: 0.01 Pa/L per second
7. Vacuum Tube: Stainless steel vacuum corrugated tube used.
8. Vacuum Sensor: Equipped with KF16 vacuum resistance gauge tube.
9. Vacuum Setting Method: The vacuum level can be freely set as needed, such as for a vacuum level of (100Pa), the gauge is set to (1.0 E 2), and for a vacuum level of (10Pa), the gauge is set to (1.0 E 1), etc. It employs upper and lower limit controls for vacuum extraction and stoppage.
10. Vacuum Control Valve: Utilizes KF40 fitting high vacuum differential pressure valve with charging pressure.
11. Vacuum Pump Configuration: Dual-stage direct-connected oil pump used.
12. Vacuum Degree Display: Utilizes a resistance vacuum gauge to display the vacuum level, with the display in scientific notation format.
V. Nitrogen Filling Device for High-Temperature Test Chambers
1. Pressure Regulator: Controls nitrogen pressure, typically set to 2-3 kg under normal operation.
2. Flowmeter: Controls the amount of nitrogen entering the box, with a range of 10L/min to 60L/min.
3. Nitrogen Filled Pipeline: Nitrogen pipeline enters the box at the very back, with an external 2-meter-long 8mm air tube.
4. Nitrogen filling port: One nitrogen port fills the box.
5. Nitrogen Source: The user company will provide the nitrogen source.
Six, High-Temperature Test Chamber Power Supply:
Voltage: AC 380V/50Hz, three-phase five-wire system
2. Current: 14A
3. Power: 9.5 kW
Section 7: High-Temperature Test Chamber Protection System
Over-temperature Protection System: Automatically shuts off heating power when the temperature exceeds the over-temperature protection setting, thereby ensuring the safety of both the product and the machine.
2. Circuit Protection Systems: Grounding protection, express insurance, phase loss protection, overload protection, circuit breakers, etc.