Shenzhen ZhongNuo Testing Technology Co., Ltd. specializes in the calibration, certification, and metrological services for gas alarms, flammable gas detectors, toxic and harmful gas detectors, and other equipment in various locations. We provide services nationwide, addressing all types of electrical safety and explosion-proof fire prevention issues for our clients, and contributing to safe production.
Power-on inspection
After powering on the instrument, it should operate normally with clear and complete display. Check the alarm function and the alarm action values.
Alarm function and alarm action value verification
The instrument's acoustic and visual alarms should be functioning normally. Insulation resistance
Insulation resistance for equipment using alternating current power should be no less than 20MΩ.
What factors can affect the calibration cycle of a combustible gas detector alarm?
Manufacturer Requirements: Flammable gas detectors from different manufacturers may have varying inspection cycle requirements. Therefore, when purchasing equipment, carefully review the product manual to understand the specific requirements of the manufacturer.
Equipment Usage Environment: The operating environment of the alarm directly affects its inspection cycle. If the equipment is frequently exposed to high concentrations of harmful gases, high or low humidity, or high or low temperatures, its inspection cycle may be shortened accordingly.
Equipment Maintenance Status: The condition of equipment maintenance is also a significant factor affecting the inspection cycle. If the equipment faces issues like battery depletion or sensor aging over a long period, the inspection cycle will accordingly be shortened.
Regular Inspection Content for Gas Detectors
Installation spacing of inspection (probe) points
2. Installation Height of the Detector
3. Indicated Error (Concentration)
4. Repeatability (Concentration)
5. Alarm Error (Concentration)
6. Response Time (Seconds)
Gas Alarm Meters Calibration Conditions and Requirements
Environmental Conditions
Ambient Temperature: 0~40℃.
2. Relative Humidity: ≤85%
3. The working environment should not interfere with the normal operation of the instrument by electromagnetic fields or interfering gases.
Calibration Tools and Instruments
1) Certified Reference Gases
The relative expanded uncertainty of the gas standard material does not exceed 2% (k=2).
2) Standard Gas Dilution Unit: Dilution Error of 1%.
3) Stopwatch: Resolution not greater than 0.1s.
4) Flow Control: Flow range not less than 500 mL/min, flow grade not less than 4.
5) Zero Point Gas: Purified compressed air.
6) Pressure Reducing Valve: A pressure reducing valve that matches with the calibrated cylinder.
7) Gas Pipeline: Tubing made of materials such as polytetrafluoroethylene or stainless steel that do not affect the concentration of the measured gas.
Shenzhen ZhongNuo Testing Technology Co., Ltd. is a third-party institution specializing in electrical safety, electrical explosion-proof services, regular certification, metrology, and calibration of gas detectors and toxic gas alarm instruments. We are licensed to address fire and explosion prevention issues for our clients.