详情描述

Shenzhen ZhongNuo Testing Technology Co., Ltd. specializes in the calibration, verification, and metrological services for gas alarms, combustible gas detectors, toxic and harmful gas detectors, and other equipment at various locations. We hold the necessary qualifications to serve customers nationwide, addressing electrical safety and explosion-proof fire prevention issues, and contributing to safe production.
Regular inspection content for gas alarms
Installation spacing of inspection (probe) points
2. Installation Height of the Detector
3. Indicated Error (Concentration)
4. Repeatability (Concentration)
5. Alarm Error (Concentration)
6. Response Time (Seconds)
淮安可燃气体探测器检测办理
The calibration cycle for flammable gas detectors is typically once a year.
In accordance with the standard JJG 693-2011 "Flammable Gas Detectors," the calibration cycle for flammable gas detectors typically does not exceed one year. This means that under normal circumstances, a comprehensive calibration and testing of the flammable gas detectors should be conducted annually to ensure their performance meets the relevant requirements.
淮安可燃气体探测器检测办理
Gas Alarm Meters Calibration Conditions and Requirements
Environmental Conditions
Ambient Temperature: 0~40℃.
Relative humidity: ≤85%.
3. The working environment should not affect the normal operation of the instrument with electromagnetic fields and interfering gases.
Calibration Tools and Instruments
1) Certified Reference Materials for Gases
The relative expanded uncertainty of the gas standard substance does not exceed 2% (k=2).
2) Standard Gas Dilution Unit: Dilution error of 1%.
3) Stopwatch: Resolution not greater than 0.1s.
4) Flow Control: Flow range not less than 500 mL/min, flow grade not less than 4.
5) Zero Point Gas: Purified compressed air.
6) Pressure Reducing Valve: A pressure reducing valve that matches with the calibrated cylinder.
7) Gas Pipeline: Tube made of materials like polytetrafluoroethylene or stainless steel that do not affect the concentration of the measured gas.
淮安可燃气体探测器检测办理
GB 12358-2024 "General Technical Requirements for Environmental Gas Detection and Alarm Instruments in Workplaces" is an important standard.
Intended to standardize the performance, functions, and testing methods of environmental gas detection alarm instruments used in work sites.
The standard was issued in 1990 and has undergone several revisions.
The new version is GB 12358-2024, which will be officially implemented on June 1, 2025.
This standard applies to the production and use of detection and alarm instruments for flammable gases, toxic and harmful gases, and oxygen in workplace environments.
Shenzhen ZhongNuo Testing Technology Co., Ltd. is a third-party service provider specializing in electrical safety, electrical explosion-proofing, gas alarms, and toxic and harmful gas detection instruments, offering regular inspections, metrological measurements, and calibrations. We are licensed to address fire and explosion prevention issues for clients.