Shenzhen ZhongNuo Testing Technology Co., Ltd. specializes in the certification, calibration, and metrological services for gas detectors, flammable gas detectors, toxic and harmful gas detectors, and other equipment in various locations. We are licensed to provide nationwide services, addressing all types of electrical safety and explosion-proof fire prevention issues for our clients, and contributing to safe production.
GB 12358-2024 "General Technical Requirements for Environmental Gas Detection and Alarm Instruments in Workplace" is an important standard.
Intended to standardize the performance, functions, and testing methods of environmental gas detection alarm instruments used in work场所.
The standard was issued in 1990 and has undergone multiple revisions.
The new version is GB 12358—2024, which will be officially implemented on June 1, 2025.
This standard applies to the production and use of gas detection and alarm instruments for flammable, toxic, and harmful gases in workplaces.
苏州气体探头检测和计量咨询
The calibration cycle for flammable gas detectors is typically once a year.
In accordance with the standard JJG 693-2011 "Flammable Gas Detectors," the calibration cycle for flammable gas detectors generally does not exceed one year. This means that under normal circumstances, a comprehensive calibration and test of the flammable gas detector is required annually to ensure its performance meets the relevant requirements.
苏州气体探头检测和计量咨询
Gas Alarm Meters Calibration Conditions and Requirements
Environmental Conditions
Ambient Temperature: 0~40℃.
2. Relative Humidity: ≤85%
3. The working environment should not interfere with the normal operation of the instrument due to electromagnetic fields and interfering gases.
Inspection Gauges and Instruments
1) Gas Standard Material
The relative expanded uncertainty of the gas standard material does not exceed 2% (k=2).
2) Standard Gas Dilution Unit: Dilution Error of 1%.
3) Stopwatch: Graduation not greater than 0.1s.
4) Flow Control: Flow range not less than 500 mL/min, flow grade not less than 4.
5) Zero Point Gas: Purified compressed air.
6) Pressure Reducing Valve: A pressure reducing valve that is compatible with the calibrated cylinder.
7) Gas Pipeline:气管, made of materials like polytetrafluoroethylene or stainless steel, which do not affect the concentration of the tested gas.
苏州气体探头检测和计量咨询
Appearance and structure
The instrument should not have any appearance damage that affects its normal operation. The surface of newly manufactured instruments should be smooth and clean, with even paint and plating layers, free from peeling and rusting.
The instrument's connections are reliable, and all dials or buttons should operate and control normally.
Logos and branding
Instrument name, model, manufacturer's name, date of manufacture, serial number, explosion-proof markings and numbers, as well as the domestic instrument's manufacturing metrological instrument license mark and number, should be clear.
Shenzhen Zhongnuo Testing Technology Co., Ltd. is a third-party institution specializing in electrical safety, electrical explosion-proofing, gas alarms, and toxic and harmful gas alarm instruments for regular inspections, metrological measurements, and calibrations. We are licensed to provide solutions for fire and explosion prevention issues for our clients.