详情描述
Shenzhen ZhongNuo Testing Technology Co., Ltd. specializes in the calibration, verification, and measurement services for gas alarms, flammable gas detectors, toxic and harmful gas detectors, and other equipment at various locations. We hold the necessary qualifications to serve nationwide, addressing a range of electrical safety and explosion-proof fire prevention issues for our clients, and contributing to safe production.
What are the contents of a calibration report for a combustible gas detector?
The verification results should be reflected in the verification certificate or report, and the verification proof or report must include the following information.
① Title, such as "Calibration Certificate" or "Calibration Report";
② Laboratory Name and Address;
③ Calibration Location (if not calibrated in the laboratory)
④Standing certificate or report (e.g., with serial number), and indicate each page and the total number of pages.
⑤Shipper's name and address
⑥ Describe and clearly identify the target school;
Calibration Date and Receipt Date
⑧ Technical specifications for calibration foundation, including name and code;
Description of the traceability and effectiveness of the measurement standards used for calibration.
⑩ Description of Calibration Environment
11. Calibration Results and Measurement Uncertainty Statement
12. Signature, title, or equivalent symbol of the issuer of the calibration certificate or calibration report, and the issuance date;
13. Calibration results are only applicable to the calibration object.
14. No part of the certificate or report may be copied without the written consent of the laboratory.
Reprüfung Intervals
Instrument recalibration is typically every 1 year, determined by the submitting unit based on actual usage. During the secondary calibration process, if there are any doubts about the instrument's test data, or if the main components of the instrument are replaced, as well as after repairs, recalibration is required.
The calibration cycle for flammable gas detectors is typically once a year.
According to the provisions of the standard JJG 693-2011 "Flammable Gas Detectors," the calibration cycle for flammable gas detectors generally does not exceed 1 year. This means that under normal circumstances, a comprehensive calibration and testing of the flammable gas detector is required annually to ensure that its performance indicators meet the relevant requirements.
The detection principle of the instruments mainly includes catalytic combustion, infrared absorption, and thermal conductivity types. The sampling methods include diffusion and aspiration. The instruments are primarily composed of detection elements, amplification circuits, alarm systems, and displays, used for monitoring the concentration of combustible gases in the environment.
Appearance and structure
The instrument should not have any appearance damage that affects its normal operation. The surface of newly manufactured instruments should be smooth and glossy, with even paint and plating coatings, and no peeling or rusting.
The instrument's connections are reliable, and all dials or buttons should operate and control normally.
Logos and Branding
Instrument name, model, manufacturer's name, date of manufacture, serial number, explosion-proof marking and number, as well as the manufacturing metrological instrument license marking and number for domestic instruments should be clear.
Shenzhen Zhongnuo Testing Technology Co., Ltd. is a third-party service provider specializing in electrical safety, electrical explosion-proofing, gas alarm systems, and toxic and harmful gas detectors, offering regular inspections, metrological services, and calibration. We are qualified to help clients address fire and explosion prevention issues.