Shenzhen ZhongNuo Testing Technology Co., Ltd. specializes in the calibration, verification, and measurement services for gas alarms, flammable gas detectors, toxic and harmful gas detectors, and other equipment in various locations. We hold qualifications to serve nationwide, addressing clients' electrical safety and explosion-proof fire prevention issues, and contributing to safe production.
Appearance and structure
The instrument should not have any appearance damage that affects its normal operation. The surface of newly manufactured instruments should be smooth and clean, with a uniform paint and coating, and no peeling or rusting.
The instrument's connections are reliable; all dials or buttons should operate and control normally.
Logos and Branding
Instrument name, model, manufacturer's name, manufacturing date, serial number, explosion-proof marking and number, as well as the manufacturing metrological instrument license marking and number for domestic instruments should be clear.
How often should various gas detectors be calibrated and checked?
The recalibration period for instruments is typically one year, which is determined by the submitting unit based on actual usage. During the secondary calibration process, if there are any doubts about the instrument test data, or if the main components of the instrument are replaced, as well as after repairs, a re-calibration is required.
Gas alarm regular inspection content
Installation spacing of inspection points
2. Installation Height of the Detector
3. Indication Error (Concentration)
4. Repetition (Concentration)
5. Alarm Error (Concentration)
6. Response Time (Seconds)
Calibration Method
Appearance and structure
Inspect visually and by touch according to the requirements of Section 4.1.
Logos and Branding
Inspect in accordance with the requirements of Section 4.2.
Power-on inspection
Inspect visually and by touch, checking the 5.3.4 alarm function and alarm action values as per the 4.3 requirements.
Introduce a standard gas substance with a concentration exceeding the alarm set point, causing the instrument to trigger an alarm. Observe whether the instrument's sound and light alarm functions normally and record the instrument's reading when the alarm is triggered. Repeat the measurement three times, with the arithmetic mean of the three readings being the instrument's alarm trigger value.
Shenzhen Zhongnuo Testing Technology Co., Ltd. is a third-party institution that provides regular calibration, measurement, and calibration services for electrical safety, electrical explosion-proofing, gas alarms, and toxic and harmful gas alarm instruments. We hold the necessary qualifications to help customers address fire and explosion prevention issues.