Shenzhen Zhongnuo Testing Technology Co., Ltd. specializes in the calibration, verification, and measurement services for gas alarms, flammable gas detectors, and toxic gas analyzers for various locations. We hold qualifications to serve nationwide, addressing clients' electrical safety and explosion-proof fire prevention concerns, and contributing to safe production.
Appearance and structure
The instrument should not have any appearance damage that affects its normal operation. The surface of newly manufactured instruments should be smooth and glossy, with a uniform paint and coating, and no peeling or rusting.
The instrument's connections are reliable, and all dials or buttons should operate and control normally.
Logos and branding
Instrument name, model, manufacturer, date of manufacture, serial number, explosion-proof marking and number, as well as the manufacturing metrological instrument license marking and number for domestic instruments, should be clear.
宿迁气体检测仪计量公司
What factors can affect the calibration cycle of a combustible gas detector alarm?
Manufacturer Requirements: Different manufacturers may have varying inspection cycle requirements for their flammable gas detectors. Therefore, when purchasing equipment, carefully review the product manual to understand the specific requirements of the manufacturer.
Equipment Operating Environment: The operating environment of the alarm can directly affect its inspection cycle. If the equipment is frequently exposed to high concentrations of harmful gases, high humidity, or extreme temperatures (either high or low), its inspection cycle may be shortened accordingly.
‌Equipment Maintenance Status: The condition of equipment maintenance is also a significant factor affecting the inspection cycle. If the equipment frequently faces issues such as battery depletion or aging sensors, its inspection cycle will accordingly be shortened.
宿迁气体检测仪计量公司
Gas Standard Material
The gas standard substance used should be of the same type as the gas measured by the instrument, such as methane or other gases; the standard gas can also use isopropanol, etc. If the instrument does not specify the type of gas being measured, the concentration of the gas at the set point of the alarm is approximately 60% of the full scale. The uncertainty of the expanded gas standard substance is not significant (%). The standard gas concentration should not exceed 0.1% of the flow rate indication error of the repetitive dilution device for the gas standard substance with a concentration not high. The concentration unit of the gas standard substance should be converted to match the unit of the tested instrument, and the deviation should not be greater than 0.1%.
宿迁气体检测仪计量公司
Gas Detector Alarm Calibration Standard Requirements
Regulation 5.5 of the "Regulation for the Verification of Combustible Gas Detectors" (JJG693-2011) stipulates: The verification cycle for combustible gas detection and alarm instruments: Generally, the verification cycle of the instrument should not exceed one year. If there is any suspicion regarding the instrument's measurement data, or if the instrument has been replaced with major components or repaired, it should be sent for verification promptly.
Section 3.5 of the "Regulations for the Use of Combustible Gas Detection and Alarm Systems" (SY6503-2000) stipulates: In-use combustible gas detection and alarm devices shall be calibrated periodically as per regulations.
Occupational Health and Safety Code for Toxic Gas Equipment Installation (GBZ/T223-2009) Article 7.2.1 stipulates: Calibration of detection and alarm devices shall be performed regularly according to the manufacturer's specified time and procedures.
Section 3.4 of the "Code for the Use of Combustible Gas Detectors and Alarms" (SY6503-2000) stipulates that the installation and use of combustible gas detectors and alarms shall have measurement instruments manufactured by recognized agencies, with **explosion-proof performance**.
Shenzhen Zhongnuo Testing Technology Co., Ltd. is a third-party institution specializing in electrical safety, electrical explosion-proofing, gas alarms, and toxic and harmful gas alarm instruments for regular inspections, metrological measurements, and calibration services. We provide qualifications to help customers address fire and explosion prevention issues.