Shenzhen ZhongNuo Testing Technology Co., Ltd. specializes in the calibration, verification, and measurement services for equipment such as gas alarms, combustible gas detectors, and harmful gas detection instruments in various locations. We hold qualifications to serve nationwide, addressing clients' electrical safety and explosion-proof fire prevention issues, and contributing to safe production.
What types of gas detectors require regular calibration and verification?
Fixed-install, handheld, and portable gas detectors, gas alarm instruments, toxic and harmful gas detectors/alarm systems, flammable gas detectors/alarm systems, gas analyzers, etc.
Gas Detector Alarm Calibration Specification Requirements
Section 5.5 of the "Regulation for Inspection and Verification of Combustible Gas Detectors" (JJG693-2011) stipulates: The verification cycle for combustible gas detection and alarm instruments: The verification cycle for the instrument generally should not exceed one year. Instruments with suspected measurement data, instruments that have replaced major components, or those repaired should be submitted for verification promptly.
Section 3.5 of the "Code for the Use of Combustible Gas Detection and Alarm Systems" (SY6503-2000) stipulates: The combustible gas detection and alarm systems in use shall be calibrated regularly as required.
Occupational Health and Safety Code for Toxic Gas Equipment Installation (GBZ/T223-2009) Article 7.2.1 stipulates: Calibration of detection and alarm devices shall be performed regularly according to the manufacturer's specified time and procedures.
Section 3.4 of "Code for the Use of Combustible Gas Detectors and Alarms" (SY6503-2000) stipulates: Combustible gas detectors and alarms for installation and use shall have measurement instruments manufactured by an authorized institution with recognized **measurement accuracy** and **explosion-proof performance**.
Appearance and structure
The instrument should not have any appearance damage that affects its normal operation. The surface of newly manufactured instruments should be smooth and clean, with a uniform paint and coating, free of peeling and rusting.
The instrument's connections are reliable, and all dials or buttons should operate and control normally.
Logos and Branding
Instrument name, model, manufacturer, date of manufacture, serial number, explosion-proof marking and number, as well as the manufacturing measurement instrument license marking and number for domestic instruments should be clear.
Gas alarm meter calibration conditions and requirements
Environmental Conditions
Ambient Temperature: 0~40℃
2. Relative Humidity: ≤85%
3. The working environment should not affect the normal operation of the instrument with electromagnetic fields and interfering gases.
Inspection gauges and instruments
1) Certified Reference Materials for Gases
The relative expanded uncertainty of the gas standard material does not exceed 2% (k=2).
2) Standard Gas Dilution Unit: Dilution Error of 1%.
3) Stopwatch: Resolution not greater than 0.1s.
4) Flow Control: Flow range not less than 500 mL/min, flow grade not less than level 4.
5) Zero Point Gas: Purified compressed air.
6) Pressure Relief Valve: A pressure relief valve that is compatible with the calibrated cylinder.
7) Gas Pipeline: Tubing made of materials like polytetrafluoroethylene or stainless steel that do not affect the concentration of the measured gas.
Shenzhen ZhongNuo Testing Technology Co., Ltd. is a third-party institution specializing in electrical safety, electrical explosion-proof, gas alarm, and toxic gas detection equipment for regular verification, calibration, and measurement services,资质齐全,dedicated to resolving fire and explosion prevention issues for clients.