Shenzhen ZhongNuo Testing Technology Co., Ltd. specializes in the calibration, certification, and metrological services for gas alarms, flammable gas detectors, toxic and harmful gas detectors, and other equipment at various locations. We offer services nationwide, addressing various electrical safety and explosion-proof fire prevention issues for clients, and contributing to safe production.
Regulatory Testing and Methods for Combustible Gas Detectors
Annual inspections are conducted on the sensors of commercial and industrial alarms, primarily checking for the alarm's indication error, repeatability, drift, and response time to ensure compliance with standards. Additionally, the suitability of the alarm's installation location and any interference from other equipment are assessed.
常州有毒有害气体探头检定
Primary Technical Requirements for Regular Calibration of Combustible Gas Detectors (Shenzhen ZhongNuo Inspection Technology Co., Ltd.)
1. Power-On Self-Check Function: The instrument should have an automatic self-check function upon startup, with a manual self-check function for alarm components as well.
2. Specify the effective detection distance: The effective detection distance of the non-contact gas alarm instrument shall not be less than 20 m.
3. Alarm Value Settings: The standard specifies specific alarm value settings for different gases.
For example, the oxygen deficiency alarm setpoint is 19.5% VOL, and the high-oxygen alarm setpoint is 23.5% VOL (refer to GB 8958—2006 "Operation规程 for Oxygen Deficiency Hazardous Work" and GB 30871—2022 "Special Operation Regulations for Hazardous Chemical Enterprises").
The first-level alarm setting for flammable gases does not exceed 25% LEL, and the second-level alarm setting does not exceed 50% LEL.
The first and second-level alarm settings for toxic and harmful gases are determined based on the OEL (Occupational Exposure Limit) or IDLH (Immediate Danger to Life and Health).
4. Error and Repeatability: The standard specifies clear requirements for the indicating error, drift, and repeatability of the instrument. For example, the indicating error for flammable gases, toxic and harmful gases, and oxygen should not exceed 5% FS, 5% FS, and 2% FS respectively within the normal test range of the instrument's full scale.
5. Calibration and Interface: The instrument should be easy to calibrate, with established calibration programs and corresponding interfaces to ensure the accuracy of its measurement results.
6. Explosion-proof Performance: Equipment intended for explosive environments must possess explosion-proof performance in compliance with relevant regulations and obtain an explosion-proof certification.
7. Audio-Visual Alarm Device: The audio-visual alarm device of the alarm should meet the visibility and sound pressure level requirements under specific conditions, ensuring effective alerting of staff during emergencies.
8. Output Control Function: Instruments with output control functionality should be capable of activating the output control function upon emitting an alarm signal, such as cutting off power supply or initiating the exhaust system, etc.
9. Power Requirements: Specific requirements are outlined for power supply voltage, continuous operating time, etc., for battery-powered and non-battery-powered instruments.
10. Other Performance Requirements: Including specifications for azimuth, response time, full-scale indication error, high-speed airflow, long-term stability, insulation withstand voltage, temperature, constant humidity and heat, electromagnetic interference resistance, steam interference resistance, vibration and drop resistance, and functionality.
常州有毒有害气体探头检定
What factors can affect the calibration cycle of a flammable gas detector alarm?
Manufacturer Requirements: Different manufacturers may have varying inspection cycle requirements for their flammable gas detectors. Therefore, when purchasing equipment, carefully review the product manual to understand the specific requirements of the manufacturer.
Equipment Operating Environment: The operating environment of the alarm can directly affect its inspection cycle. If the equipment is exposed to high concentrations of harmful gases, high or low humidity, or high or low temperatures for extended periods, its inspection cycle may be shortened accordingly.
‌Equipment Maintenance Status: The condition of equipment maintenance is also a significant factor affecting the inspection cycle. If the equipment faces issues like battery depletion or sensor aging over a long period, the inspection cycle will accordingly be shortened.
常州有毒有害气体探头检定
Routine inspection items for workplace environmental gas detection and alarm instruments:
Alarm sound and light (if applicable)
2. Low response value
3. Indication Error (for Gas Detectors/ Gas Detector Alarms)
4. Instrument Drift (in Portable Instruments)
5. Alarm Error (for Gas Detection Alarm/ Gas Alarm)
6. Pump-actuated instrument gas collection distance and flow rate (for pump-actuated instruments)
7. Power Supply (for battery-powered instruments)
8. Response Time
9. Status Indicator (with wireless communication capability)
Shenzhen ZhongNuo Testing Technology Co., Ltd. is a third-party institution specializing in electrical safety, electrical explosion-proofing, gas detectors, and toxic and harmful gas alarm instruments for regular verification, measurement, and calibration services. We hold the necessary qualifications to address fire and explosion prevention issues for our clients.