PL-1000FProduct Technical Features: Dual plunger parallel pump head design, solvent compression compensation, multi-point(Multiple10Dot)Flow curve calibration for high-precision fluid transportation across the full range of flow rates. Low pressure pulsation, utilizing cam curve compensation and flow pulsation electronic technology for effective fluid pressure pulsation control; floating low-pressure seal design reduces radial stress on high-pressure seals, extends seal life, and offers high reliability.4.3TFT color screen display, intuitive and easy operation; control interfaces includeRS-232/RS-485Interface, capable of computer reverse control; supports multiple communication protocols: including,Modbus 、SSICompatibility protocols, accessible viaClarityChromatography software control
PL-1000FProduct Technical Features:
Dual plunger tandem pump head design, solvent compression compensation, multi-point(Multiple10dot)Flow curve calibration for high-precision fluid transport across the full range of measurement.
Low pressure pulsation, utilizing cam curve compensation and flow pulsation electronic technology, effectively controlling fluid pressure pulsation.
Floating low-pressure seal design, reduces radial stress on high-pressure seals, extends seal lifespan, high reliability.
4.3Inch color screen display, intuitive and convenient operation;
Control interfaces include:RS-232/RS-485Interface, capable of computer reverse control; supports various communication protocols: including,Modbus 、SSICompatibility Protocols, available forClarityChromatography software control
Project | Index | |
1 | Infusion Method | Instrument Features:1Continuous Conveyance;2Accurate traffic; |
2 | Flow Range | 0.01-1000.00/min |
3 | Incremental | 0.01ml/min |
4 | Flow Accuracy | ± 1% |
5 | Repeat Traffic | <±0.5% |
6 | Pressure Range | 0-15Mpa |
7 | Pulse Pressure | ≤ 0.1Mpa |
8 | Flow Path Materials | 316LStainless steel, Ruby.PTFECeramics |
9 | Pipe linkage | 1/16"Standard Pipeline Link |
10 | Display Parameters | LCD Display |
11 | Control Method | Manual panel control or computer reverse control |
12 | Power Supply | 220V,50Hz |
13 | Dimensions | 130×210×110mm(Length)×High×Deep/2.8Kg |
14 | System Protection | Set protection for high pressure: Pressure reaches the set upper limit, alarms and stops operation. ; Recover range pressure, pump auto-activates. |