品牌: Xiangyuan Instruments
Brand: Xiangyuan Instruments
Applicable Medium: Tap water
Installation Direction: Horizontal installation
单价: 440.00/Piece
最小起订Quantity: 1 Piece
供货总Quantity: 10000 Piece
有效期至: 长期有效
最后更新: 2022-11-16 15:20
 
详细Info

The ultrasonic heat meter is composed of a flow sensor, a matching temperature sensor, and a calculator, featuring a compact structure and easy installation.

Our products utilize ceramic piezoelectric transducers, ensuring high accuracy and stability; with no mechanical movement, no wear, and resistance to poor water quality, as well as low maintenance costs; they can be installed horizontally or vertically, and can be placed on the inlet or return piping according to user needs (preselection required); RS485 interfaces can be added upon customer request to enable remote meter reading functions according to different requirements, facilitating centralized management.

 

Nominal Bore Size (mm)

Traffic

Common Traffic

Traffic

Table body length

m3/h

m3/h

m3/h

mm

DN15

0.03

1.5

3

110

DN20

0.05

2.5

5

130

DN25

0.07

3.5

7

160

DN32

0.12

6

12

180

DN40

0.2

10

20

200

DN50

0.6

15

30

200

DN65

1

25

50

200

DN80

1.6

40

80

225

DN100

2.4

60

120

250

DN125

4

100

200

250

DN150

6

150

300

300

DN200

10

250

500

350

DN250

16

400

800

450

DN300

24

600

1200

500

Water Flow Reading (m³)

999999.99

Heat Reading (kW·h)

99999999

Accuracy Grade

2nd Grade

Pressure loss

<25 kPa (at operating flow)

Work pressure

1.6MPa

Heat consumption calculation

Starting from 0.25K

Temperature Range

(4~95)℃

Temperature range

(3~75)K

Temperature Resolution

0.01℃

Environmental Grade

Type A (5-55)°C

Battery Operating Time

≥6 years (Lithium-ion Batteries)

Temperature Sensor

PT1000 Platinum Resistance

Communication Mode

Infrared; M-Bus; RS-485 (optional)

The heat meter must be installed in a dry, accessible location.

2. Install the heat meter using professional installation tools.

3. The heat meter must be installed according to the instructions on the label (horizontal/vertical), with the text facing up, and the arrow direction on the meter body (including the filter) aligned with the flow direction of the HVAC piping. The outlet should be higher than the heat meter; when installed vertically, the flow direction must be upwards.

4. Ensure that new piping is thoroughly cleaned of debris such as stones, sand, and fibers before installing the heat meter to prevent malfunctions; according to the individual heating metering design requirements, a filter must be installed before the heat meter's inlet; the filter must be cleaned and maintained regularly.

5. To achieve measurement accuracy, leave a straight pipe section of approximately 10D length at the heat meter's inlet, where D is the heat meter's pipe diameter. The straight section at the front end of the outlet should not be less than 5 times the pipe diameter. When installed at the confluence of two return water pipelines, the heat meter's connection head (such as a T-joint) requires a straight pipe section of 10 times the pipe diameter to ensure uniform mixing of water temperatures between the two pipelines.

6. Data is collected from the heat meter via M-Bus bus. Do not connect this wire to the municipal power supply; otherwise, it may damage the meter.

7. The heat meter has a 485 interface at the top row; please note the polarity of the A and B lines when wiring.



询价单
主题 *
内容 *
Your company's名
Contact person *
ContactPhone *
电子Email
验证码  
 点击确定代表您同意《服务条款》《隐私政策》
 
 Click submit means you agree to《Service terms》《Privacy policy》

13933067893