详情描述

Direct measurement results derivation: The oxygen probe and its transmitter measure and output two key signals: EMF (unit: mV) and zirconium tube temperature T (unit: °C or K). Calculating oxygen partial pressure: Using the Nernst equation to infer the oxygen partial pressure P_sample in the measured gas/environment: P_sample = P_ref / exp(EMF * 4F / (RT))