I. Applications of High and Low Temperature Humidity Test Chambers
This equipment is primarily designed for testing the adaptability of electrical equipment, electronic products, their components, and other materials under high and low temperatures, as well as humid and wet conditions during storage, transportation, and use. The test equipment is mainly used to simulate environmental conditions for products according to national standards or customer-defined requirements, assessing the physical and other related properties of the products under high and low temperature conditions. After testing, the performance of the products is evaluated through inspections to determine if they still meet the specified requirements for product design, improvement, certification, and factory inspection.
Section II: Key Technical Specifications of the Equipment
Temperature Range: -20℃ to -80℃; +150℃
2. Humidity Range: 20~98% R.H.
3. Temperature fluctuation: ±0.5℃
4. Temperature Deviation: ≤2℃
5. Cooling Rate: 0.7-1℃/min
Section 3: Reference Specifications
| Model | Studio Size (mm) (W x D x H) | Case Size (mm) (W x D x H) |
| HS-50 | 320×350×450 | 800×840×1560 |
| HS-100 | 400×500×500 | 900×950×1610 |
| HS-150 | 500×500×600 | 950×1000×1710 |
| HS-225 | 500×600×750 | 1003×1050×1860 |
| HS-250 | 500×600×810 | 1060×1080×1890 |
| HS-500 | 700×800×900 | 1250×1200×1910 |
| HS-800 | 800×1000×1000 | 1250×1500×2010 |
| HS-010 | 1000×1000×1000 | 1450×1500×2010 |
Note: Customized non-standard sizes available based on customer's technical requirements!
Section 4: Structural Characteristics of the Equipment
1. The equipment is mainly composed of the housing, refrigeration system, air circulation system, and control system.
2. The housing is made of cold-rolled steel plates with double-sided electrostatic spraying, and the inner lining is constructed from SUS304 stainless steel plates. The door features a large observation window in the center, equipped with an observation light, allowing users to clearly view the test conditions of the sample.
3. The overall appearance is aesthetically pleasing and generous. The thermal insulation layer is ultra-fine glass wool, offering high strength and excellent thermal insulation properties.
4. The equipment primarily uses an intelligent digital temperature and humidity controller, which is easy to learn and operate, and different functional levels of instruments are interoperable. The input features a digital calibration system, with built-in non-linear correction tables for common thermocouples and thermistors, ensuring stable measurements. Equipped with on/off and PID artificial intelligence adjustment functions, with 0.1 grade accuracy and multiple alarm modes. Features independent heating and cooling, and a unique BTHC balanced temperature control method.
5. The refrigeration system features a fully enclosed imported compressor set from Europe's Tecumseh, with a mechanical single-stage refrigeration or cascade low-temperature circuit system, and an automatic control and safety protection coordination system.
6. Humidity is circulated by evaporative means using a surface of stainless steel humidifier.
7. Electrical appliances are executed using Schneider's product series, ensuring safety and stability.
V. Relevant Standards
GB 10589-89 Technical Conditions for High and Low Temperature Test Chambers
GB/T5170.2-1996 Basic Method for Calibration of Fundamental Parameters of Environmental Test Equipment for Electrical and Electronic Products - Temperature Test Equipment
GB2423.1-89 Basic Test Procedures for Electrical and Electronic Products: Test A - Low Temperature Test Method
GB/T2423.1-2001 GB/T2423.2-2001 GB/T2423.1993 GB/T2423.4-1993
Section 6: Safety Protection Devices for Equipment
Overload, short-circuit protection, and leakage protection.
2. Grounding Protection
Over-temperature Protection
4. Over-pressure and Overload Protection
5. To protect the equipment, all alarms will automatically cut off the power and emit audio alerts.
6. Humidity and Water Shortage Protection
Section 7: Conditions at the Equipment Usage Site
Temperature: 15℃~35℃
2. Relative Humidity: Not greater than 85% RH
3. No intense vibration or electromagnetic field interference nearby
4. No high concentration of dust or corrosive substances in the vicinity.
5. No direct sunlight or other heat sources radiation
6. Surrounding air currents should be mild, and when forced air is required, it should not blow directly onto the case.
7. The testing box should be placed flat and kept level.
8. There should be sufficient space around the test box for easy maintenance and operation.
9. The installation site has good ventilation
10. Good Grounding
Section 8: Service Commitment
1. Free on-site installation, debugging, and operator training.
2. The product comes with a one-year warranty from the date of delivery to the user, except for the following circumstances:
a. Buyer's operation error resulted in human error and damage.
b. Other circumstances not to be borne by the supplier
3. Offer paid services after the warranty period expires.