I. Applications of the Intelligent Constant Temperature and Humidity Test Chamber
This equipment is primarily designed for testing the adaptability of electrical equipment, electronic products, their components, and other materials under high and low temperatures, as well as in humid and cyclic conditions during storage, transportation, and use. The test equipment is mainly used to simulate environmental conditions according to national standards or customer-defined requirements, and to assess the physical and other related properties of products under alternating high, low temperature, and humidity cycles. After testing, the performance of the products is evaluated through inspections to determine if they still meet the specified requirements, for product design, improvement, certification, and factory inspection purposes.
II. Main Technical Specifications of the Equipment
Temperature Range: -20°C, -40°C, -50°C, -60°C, -70°C, -80°C to +150°C
2. Humidity Range: 20~98% R.H.
3. Temperature fluctuation: ±0.5℃
4. Temperature Deviation: ≤2℃
5. Cooling Rate: 0.7-1℃/min
Section 3: Reference Specifications (ZHS/GDJS)
Model Studio Size (mm)
(W·D·H) Case Dimensions (mm)
(W·D·H)
ZHS-50 320×350×450 800×840×1560
ZHS-100 400×500×500 900×950×1610
ZHS-150 500×500×600 950×1000×1710
ZHS-225 500×600×750 1003×1050×1860
ZHS-250 500×600×810 1060×1080×1890
ZHS-500 700×800×900 1250×1200×1910
ZHS-800 800×1000×1000 1250×1500×2010
ZHS-010 1000×1000×1000 1450×1500×2010
Note: Custom non-standard sizes available according to customer's technical requirements!
IV. Structural Characteristics of the Equipment
1. The equipment is mainly composed of the housing, refrigeration system, air circulation system, and control system.
2. The cabinet shell is made of cold-rolled steel plates with double-sided electrostatic spraying, the inner lining is made of SUS304 stainless steel plate. The door is equipped with a large observation window in the middle, complete with an observation light, allowing users to clearly see the test condition of the sample.
3. The overall appearance is elegant and tasteful. The insulation layer is ultra-fine glass wool, featuring high strength and excellent thermal insulation properties.
4. Equipped with a human-machine dialogue operation mode, an imported full-color touch screen temperature and humidity controller. It achieves automatic operation of set values, cycles, and programs. The curve of operating data can be automatically saved or connected to a computer for remote control. Features on/off control and AI artificial intelligence adjustment functions, with 0.2-grade accuracy and various alarm modes. Capable of heating, cooling, humidity control, and maintaining constant temperature and humidity, with a unique BTHC balanced temperature adjustment method.
5. The refrigeration system features a fully enclosed imported compressor set from Europe's Tecumseh, with a mechanical single-stage refrigeration or cascade low-temperature circuit system, and an automatic control and safety protection coordination system.
6. Humidity is circulated using a stainless steel humidifier with surface evaporation.
7. Electrical equipment is executed using Schneider's product series, ensuring safety and stability.
V. Relevant Standards
GB 10589-89 Technical Conditions for High and Low Temperature Test Chambers
GB/T5170.2-1996 Method for Verification of Basic Parameters of Environmental Test Equipment for Electrical and Electronic Products - Temperature Test Equipment
GB2423.1-89 Basic Test Procedures for Electrical and Electronic Products - Test A: Low Temperature Test Method
GB/T2423.1-2001 GB/T2423.2-2001 GB/T2423.1993 GB/T2423.4-1993
Section 6: Safety Protection Devices for Equipment
Overload, short-circuit protection, and leakage protection.
2. Grounding Protection
Over-temperature Protection
4. Overpressure and Overload Protection
5. To protect the equipment, all alarms will automatically shut off the power and emit audio alerts.
6. Humidity Shortage Protection
Section 7: Conditions at the Equipment Usage Site
Temperature: 15℃~35℃
2. Relative Humidity: Not greater than 85% RH
3. No severe vibration or strong electromagnetic field interference in the vicinity
4. No high concentration of dust or corrosive substances in the vicinity.
5. No direct sunlight or other direct heat sources
6. No strong air currents should be present around the unit, and when forced airflow is required, the air should not blow directly onto the box.
7. The testing box should be placed flat and kept level.
8. Allow sufficient space around the test box for easy maintenance operations.
9. The installation site has good ventilation
10. Good Grounding
Section 8: Service Commitment
1. Free on-site installation, debugging, and training for operators.
2. The product comes with a one-year warranty from the date of delivery to the user, except for the following situations:
a. Buyer's operation error resulted in human-induced faults and damage.
b. Other circumstances not to be borne by the supplier
3. Post-warranty paid services are provided.