I. Application and Usage Scope
This product is suitable for measuring formaldehyde emissions from various artificial boards, composite wood flooring, carpet, carpet padding, and carpet adhesives, as well as for the isothermal and isohyetal treatment of wood or artificial boards. It can also be used for detecting volatile harmful gases in other building materials. The max feature of this product lies in its ability to simulate indoor climate conditions to the maximum extent, resulting in test results that are closer to real-world environments, thereby ensuring authenticity and reliability.
This product complies with the following standards:
"Formaldehyde Emission from Wood-based Panels and Their Products" GB18580
"Limit of Hazardous Substances in Wood Furniture" GB18584
"Limit of Hazardous Substance Emission from Indoor Decorative Materials, Carpet, Carpet Padding, and Carpet Adhesives" GB18587
"Test Methods for Physical and Chemical Properties of Engineered Wood and Laminated Wood Products" GB17657
Standard Test Method for Determining Formaldehyde Concentration in the Emitted Gases from Wood Products in Small-Scale Environmental Chambers - ASTMD6007
Model: LBZ-E1000A
III. Equipment Technical Specifications
Project | Performance Parameters | |
PrimaryTechnologyIndex | Studio size | Inner box size: 1 ± 0.02 m³ |
Humidity and Temperature Control Range | Temperature Range: +15℃ to +50℃ | |
Humidity Range: 30% to 70% RH | ||
Instrument resolution | Temperature: 0.1℃ | |
Humidity: 0.1%RH | ||
Volatility | Temperature: ≤0.5°C | |
Uniformity | Temperature: ≤1℃ | |
Humidity and temperature deviation | Temperature: (-0.5°C, 0.5°C) | |
Humidity: (-2%RH, 3%RH) | ||
Studio Airflow | The wind speed at the center of the box is 0.1 to 1 m/s, ±5% (adjustable wind speed) | |
Air exchange rate range | 0.2 to 2 times per hour | |
Flow regulation range must be greater than 0.2 to 2 m³/h, ±5%, meaning 0.2 min value, 2 non-flow min value, with a relatively large accuracy error in the extreme values. | ||
Flow meter accuracy | Glass转子流量计, accuracy not less than 2.5 grade | |
Sealability | Method 1: Overpressure of 1000pa corresponds to a leakage rate less than 10-3 x 1 m3/min. | |
Stable Duration | Stable temperature time less than 40 mins; stable humidity time less than 80 mins, equipment can continuously operate for at least 30 days. | |
Baseline Indicator | Formaldehyde background concentration: ≤0.006 mg/m³ | |
Noise | No resonance sound from the box door equipment at or below 65 decibels. | |
Other | Ensure no liquid or solid water is present within the work chamber. | |
Structural Method | The inner lining corners feature large chamfered design, with the working compartment fully welded for enhanced sealing performance. | |
CartonBodyTieStructure | Material Composition | Exterior Material: 1.5MM cold-rolled steel plate with electrostatic powder coating, color as shown in the pictures. |
Recirculating fan unit | Full metal blade axial flow fan with oil-free bearings; on the rear faceBottom side mounted with 2 exhaust fans(Blade position lowered, fan diameter edge 150MM from bottom), top porous plate exhaust, air circulation for easy disassembly and cleaning, beveled section with large chamfer, non-abrasive, easy to clean. | |
Studio | One-piece molded working cabin, seamless welding, polished weld surface, no blind spots or seams. | |
Hatch door | Hinged door, multi-point pressure seal, adjustable gap (single door) | |
Observation window | Front door centered position, 3-layer hollow glass with heating element, featuring anti-condensation function. | |
Sampling hole | Positioned on the left side of the equipment, there are 3 air sample collection points inside the cabin. The sampling ports are made of brass and can be connected with a 6x9 latex hose. | |
Lighting fixture | Utilizing LED strip lighting, designed on the exterior observation cabin, for easy observation of samples inside. | |
Air Conditioning | Supply Gas | Electromagnetic compression air pump, independent power plug design |
Temperature Control | The evaporator is located within the test chamber's air duct, controlling temperature with recirculating water. | |
Humidity Control | Desiccant Dehumidification Method: 5-40°C; Stainless steel balls and water spray humidification; filling material less than 40cm; ensure that the humidity of the exhaust gas is saturated; float balls made of stainless steel to ensure no formaldehyde release; level gauge treated for anti-corrosion. | |
Intake Volume Control | Flowmeter needle valve, intake with porous tube for even release in the working chamber | |
Air intake purification | An air pump followed by a series air purification unit (dew point bucket) | |
Refrigeration | Cooling Compressor | 2 fully enclosed compressors (imported from France, Tai康) |
Heat exchange method | Finned evaporator, installed inside the recirculating air duct. | |
Condensation Method | Air-cooled | |
Refrigerant | Environmentally Friendly Refrigerant R404a | |
Sensor | Temperature Control Sensor | Sheathed Platinum Resistance Thermometer (Import) |
Temperature Transmitter | Kunlun Coast Temperature and Humidity Sensor | |
EquipmentControl | Controller | Delta 7-inch Temperature and Humidity Controller (Color LCD Touch Screen) |
Human-Machine Dialogue Interface | Colorful 7-inch LCD Display, Touch Key Settings | |
Instruction Precision | Temperature: 0.1°C, Humidity: 0.1%RH, Time: 1s | |
Operation | The product features a resistive high-precision touch screen for operation, allowing direct use of a USB mouse or stylus to click on the screen for function selection or parameter setting. | |
Operation Mode | Automatic and manual operation | |
Display Item | Temperature, Measured Temperature, Set Humidity, Measured Humidity, Dew Point Set Temperature, Dew Point Measured Humidity, Total Operating Time, including Historical Data Query, Real-time Curve and Historical Data Playback, Fault Records | |
Established Project | Work cabin temperature, work cabin humidity, dew point temperature, start-up time | |
Record Function | Work cabin temperature, humidity, dew point temperature; equipped with USB interface, supports USB flash drive storage and backup. | |
Data Correction | Work cabin temperature, humidity, and dew point humidity are adjustable. | |
Others | Features permission level settings to prevent user errors. | |
SafetyEquipment | Work cabin | Over-temperature alarm protection |
Refrigeration Compressor | Overpressure protection for fans, overload protection | |
Grounding Protection | Safety grounding protection for structural components; insulation resistance of electrical components and equipment enclosures not less than 0.5M ohms | |
Heater | Heater Short Circuit Protection | |
Humidifier | Humidifier Low Water Protection | |
Condenser | Condenser Low Water Protection | |
Label | a) Product Model, Name; b) Manufacturer's Name, Address; d) Factory Serial Number | |
logo | Electronic Label | |
Warning Sign | Warning signs are complete | |
Other: Test chamber designed for water washdown, easy to disassemble and install air duct baffles | ||
Section 4: Equipment Design Advantages
Design Advantage 1:
The variable frequency wind speed regulator is set at the rear of the housing, designed to recessed, making it less prone to accidental contact. |
2. Place the fans at the very back, two fans, positioned near the bottom. |
3. The vent hole diameter on the wind panel is reduced to approximately 2-3mm. |
4. The bottom vent holes are modified to resemble louvers, making it impossible to see the internal compressor from the outside, while the ventilation holes on the other three sides are very small. |
5. Electrical loads are isolated from each other. |
6. The case is well-filled with cushioning material, particularly around the door, ensuring minimal noise when the door is opened or closed. |
Design Advantage 2:
The mechanical performance of the door handle assembly machine should be excellent. |
2. 3 sampling holes, with a pagoda joint at the connection of the inner and outer diameters. |
3. The caster support triangle is reinforced with a 5mm thickness. |
4. The internal air circulation blades have no protruding exhaust holes; they are internally fitted with a wind ring. |
5. The box door features a double door swing-open mechanism. |
6. The electric cabinet doors use dowel pins instead of screws, and reinforced ribs are added to the door. |
7. The internal fan positioning is accurately adjusted. |
8. Color: The equipment features a computer white appearance. |
Section 5: Equipment Accessories
1. Internal Sample Bracket: 2 pcs |
2. Instruction Manual: 1 copy |