Handle weights with care.
Before use, clean the surface dust of the weight with a soft brush or cloth.
3. Inspect the weights inside the box before and after use to prevent loss.
4. Weights should be properly stored and not placed in a damp and dusty environment.
5. Do not touch the weights with bare hands. Use a tool to remove the weights when handling.
6. Before calibrating the balance, the weights must be maintained at a constant temperature to achieve the laboratory's temperature conditions.
7. Sets of weights must not be mixed with other weights.
Avoid using thick acids, alkalis, and greases during use.
9. After use, clean the weights with alcohol and let them air dry naturally before storing them in the original packaging box for future use.
10. Weights should be calibrated regularly based on usage frequency, and calibration records should be maintained.
11. If there's an issue with the weight performance, it should be sent to a testing facility for inspection immediately.
Lock Type 10kg