Present Phenomenon | Original Principle | Processed |
1. Powerless | 1. Plug without power | Replace Power Outlets |
2. Plug not fully inserted or wire disconnected | Plug in the power cord or connect the wire. | |
3. Circuit Breaker Opened | 3. Replace the fuse | |
4. Power switch not engaged | 4. Turn off the power switch | |
2. The internal temperature of the box does not rise. | Set temperature low | Adjust settings temperature |
2. Electric heater malfunction | 2. Exchangeable Electric Heater | |
3. Temperature Controller Malfunction | 3. Temperature Controller Replacement | |
Temperature sensor connection wire loose | 4. Tighten the sensor connection wire nuts | |
5. The over-temperature switch's adjustment range is too small. | 5. Adjust to a value greater than the instrument's target value | |
3. The set temperature varies greatly from the internal temperature of the chamber. | Temperature sensor malfunction | 1. Replacing the temperature sensor |
2. Elevated ambient temperature (above 30℃) | 2. Use in air-conditioned environments | |
4. Temperature Out of Control | Temperature sensor detachment issue | Fixed Temperature Sensor |
2. Temperature controller and thyristor malfunction | 2. Replace temperature control gauges, thyristors | |
5. The spinner is not functioning. | Speed too low Insulation and thermal protection for motors | Increase speed to the right 2. Overheating occurred; reduce operating temperature. |
6. Squealing noise | Motor oil deficiency Lack of lubrication in bearings | 1. Top up promptly 2. Refuel promptly |
The Air Change Turntable Aging Test Chamber creates a stable temperature and air environment for long-term testing of batch material failure or fatigue resistance, utilizing scientific methods. It is suitable for soft materials used in automotive parts, plastic components, wire and cable parts, sealing silicone or rubber materials, etc.