I. Overview:

The hydrogen generator electrolytic cell employs the internationally advanced solid polymer electrolyte hydrogen production technology, yielding hydrogen with oxygen content less than 0.1ppm. It uses pure water as the electrolyte and titanium metal as the electrodes. After applying electricity, hydrogen is produced at the cathode and oxygen at the anode. The oxygen is vented, while the hydrogen is separated from water in the gas-liquid separator and then fully dried through a drying tube before being emitted through the exhaust port. This instrument incorporates a high-sensitivity, fuzzy automatic tracking control system, producing hydrogen on demand and stopping when not in use, achieving automatic constant pressure and flow control. This ensures a pressure stability with an accuracy range less than 0.001 Mpa, guaranteeing high purity, stable pressure, safety, and continuous hydrogen output.


Appendix Image:

 

II. Main Technical Parameters

Hydrogen Purity:

2. Hydrogen Flow Rate: 0-10 L/min

3. Output Pressure:0.4 MPa (Factory adjustable from 0-1.0 MPa)

4. Available Pure Water: Water with a resistivity greater than 2MΩ, deionized water, or distilled water for the second time

5. Pressure Stability: <0.001 Mpa

6. Power Supply: 220V ± 10%, 50Hz

7. Power Consumption: 3600W

8. Dimensions: 700mm x 450mm x 700mm

Net Weight: 45 kg


III. Causes of Failure and Solutions

Fault Phenomenon

Fault Cause

Inspection Methods | Exclusion Methods

The instrument fails to operate upon startup.

Power light is not illuminated

1. The circuit is not connected.

2. Fused switch burned out

Connect the circuit properly

Replace fuse

Hydrogen gas pressure is insufficient

Flow display approximately 10L

Instrument internal airway system leak

Check all gas line connections with a detection fluid

Tighten the leak point

Flow display is around 10L

Hydrogen gas equipment, pressure is off

Instrument external air line connection leakage

Hydrogen gas flow rate value

Excessive deviation

Electrolytic cell aging

2. Abnormal power output current

1. Replace electrolytic cells, clean electrolytic cells

2. Adjust power supply, replace power supply

Hydrogen impurity

Silica gel in the drying tube has reached saturation.

Replace color-changing silicone

There is water inside the purification and drying tube.

The air separator is unable to drain properly.

Replace air-water separator

Clean separator for gas and water

Excessive consumption of pure water

1. There is a leak in the water recirculation pipeline

2. Electrolytic cell temperature too high

Observe and eliminate leakage

Electrolytic cell aging: Replace electrolytic cell or clean electrolytic cell