- AllProduct Category
-
Deuterium Generator
Low-Pressure Nitrogen Generator
Maintenance-Free Generator Series
Liquid Nitrogen Generator
Hydrogen Air Generator
Nitrogen Oxygen Integrated Machine
Nitrogen-Hydrogen Air Generator
Pure Water Oxygen Generator
Air Generator
Vapor Phase Nitrogen Generator
Hydrogen Generator
Pure Water Type Hydrogen Generator
High-Pressure Hydrogen Generator


详情描述

I. Overview
PGD-600BDeuteriumGas Generator Electrolytic Cells Utilize the World's Most Advanced Solid Polymer Electrolyte TechnologyDeuteriumTechnology, to enhance outputDeuteriumOxygen content in the gas is less than0.1ppmIt uses pure water as the electrolyte and titanium metal as the electrodes. After being powered on, the cathode produces...DeuteriumGas, oxygen produced from the anode. Oxygen vented.DeuteriumAir enters the air-water separator and is separated from water. After being fully dried through the drying tube, it is output from the exhaust port. This instrument employs a high-sensitivity, fuzzy automatic tracking control system, producing gas on demand and stopping production when not in use, achieving automatic constant pressure and flow control, with pressure stability and accuracy range less than0.001MpaGuarantee instrument outputDeuteriumHigh purity, high pressure Stable, secure, and continuous.
Two,Key Technical Parameters
Electrolytic Cell: Pure Water Electrolytic Cell
3、DeuteriumAirflow:0-600ml/min
4Output Pressure:≥0.4MPa
5Pure Water Available: Greater than2MΩDeionized water or double-distilled waterOr Yibao Water
6Operating Temperature:-15℃-- 50℃
7Pressure Stability:<0.001Mpa
8Power Supply:220V±10% 50Hz
9Power Consumption: 180W
10Dimensions:360mm×190mm×350mm
11Bowl volume:≥2.0L
12Machine Weight: Approximately13kg
III. Causes of Failure and Troubleshooting Methods
Fault Phenomenon | Reason for Failure |
Inspection Method | Exclusion Method
|
Equipment fails to operate upon startup Power indicator light is not illuminated | 1、 The circuit is not connected. 2、 Fusible link burned out | Connect the circuit properly Replace fuse |
ProductDeuteriumPressure is not building up Flow display300Above | Instrument internal airway system leak
| Check connections with a test liquid Tighten leak points |
Flow display150Left and right Please provide the Chinese content that needs to be translated.DeuteriumGas instruments, pressure not exceeding | Instrument external air line connection leak | |
ProductDeuteriumFlow Rate Value Significant deviation | 1Electrolytic cell aging 2Abnormal output current of power supply | 1、 Replace electrolytic cells, clean electrolytic cells 2、 Adjust power supply, replace power supply |
DeuteriumImpure gas | Silicone adsorption in the drying tube has reached saturation. | Changeable color silicone |
Water inside the purification and drying tube | GasWater separator is not draining properly | Replace air-water separator Clean separator for gas and water |
Excessive pure water consumption | 1、 Leakage in the water recirculation pipeline 2、 Electrolytic cell temperature is too high | Observe and eliminate water leakage Electrolytic cell aging - Replace electrolytic cell or clean electrolytic cell |