详情描述


I. Overview

PGD-600BDeuteriumGas Generator Electrolytic Cells Utilize the World's Most Advanced Solid Polymer Electrolyte TechnologyDeuteriumTechnology, to enhance outputDeuteriumOxygen content in the gas is less than0.1ppmIt uses pure water as the electrolyte and titanium metal as the electrodes. After being powered on, the cathode produces...DeuteriumGas, oxygen produced from the anode. Oxygen vented.DeuteriumAir enters the air-water separator and is separated from water. After being fully dried through the drying tube, it is output from the exhaust port. This instrument employs a high-sensitivity, fuzzy automatic tracking control system, producing gas on demand and stopping production when not in use, achieving automatic constant pressure and flow control, with pressure stability and accuracy range less than0.001MpaGuarantee instrument outputDeuteriumHigh purity, high pressure   Stable, secure, and continuous.


Two,Key Technical Parameters

    Electrolytic Cell: Pure Water Electrolytic Cell

     3DeuteriumAirflow:0-600ml/min

4Output Pressure:0.4MPa

5Pure Water Available: Greater than2MΩDeionized water or double-distilled waterOr Yibao Water

6Operating Temperature:15℃-- 50℃

7Pressure Stability:<0.001Mpa

8Power Supply:220V±10%    50Hz

9Power Consumption: 180W

10Dimensions:360mm×190mm×350mm

11Bowl volume:2.0L

12Machine Weight: Approximately13kg

III. Causes of Failure and Troubleshooting Methods

Fault Phenomenon

Reason for Failure

 

Inspection Method | Exclusion Method

 

Equipment fails to operate upon startup

Power indicator light is not illuminated

1、 The circuit is not connected.

2、 Fusible link burned out

Connect the circuit properly

Replace fuse

ProductDeuteriumPressure is not building up

Flow display300Above

Instrument internal airway system leak

 

Check connections with a test liquid

Tighten leak points

Flow display150Left and right

Please provide the Chinese content that needs to be translated.DeuteriumGas instruments, pressure not exceeding

Instrument external air line connection leak

ProductDeuteriumFlow Rate Value

Significant deviation

1Electrolytic cell aging

2Abnormal output current of power supply

1、 Replace electrolytic cells, clean electrolytic cells

2、 Adjust power supply, replace power supply

DeuteriumImpure gas

Silicone adsorption in the drying tube has reached saturation.

Changeable color silicone

Water inside the purification and drying tube

GasWater separator is not draining properly

Replace air-water separator

Clean separator for gas and water

Excessive pure water consumption

1、 Leakage in the water recirculation pipeline

2、 Electrolytic cell temperature is too high

Observe and eliminate water leakage

Electrolytic cell aging - Replace electrolytic cell or clean electrolytic cell