详情描述

I. Overview

PGD-600BDeuterium GasThe generator's electrolytic cell utilizes the internationally advanced solid polymer electrolyte deuteration technology, enabling the production ofDeuterium gasOxygen content less than 0.1ppm, it uses pure water as the electrolyte and titanium metal as the electrode. Upon applying power, the cathode produces...Deuterium GasAnode produces oxygen. Oxygen is vented.Deuterium gasAfter entering the air-water separator, separation from water is achieved. The air is then thoroughly dried in the drying tube and released through the exhaust port. This instrument features a high-sensitivity fuzzy automatic tracking control system, producing gas on demand and only when in use, thus eliminating gas production when not in use. It achieves automatic constant pressure and flow control, ensuring that the pressure stability accuracy is less than 0.001 MPa, guaranteeing the output of the instrument.Deuterium gasHigh purity, pressure stable, safe, and consistent.


II. Main Technical Parameters

   Electrolytic Cell: PME Pure Water Electrolytic Cell

   3、Deuterium gasFlow Rate: 0-600 ml/min

4. Output Pressure: ≥0.4 MPa

5. Available Pure Water: Water with a resistivity greater than 2MΩ, distilled water, or Yibao water

6. Operating Temperature: -15℃ to 50℃

7. Pressure Stability: <0.001 Mpa

8. Power Supply: 220V ±10% 50Hz

9. Power Consumption: 180W

10. Dimensions: 360mm x 190mm x 350mm

11. Sink Volume: ≥2.0L

12. Machine Weight: Approximately 13kg

III. Causes of Failure and Troubleshooting Methods

Fault Phenomenon

Fault Cause

 

Inspection Methods | Exclusion Methods

 

Instrument fails to operate upon startup

Power indicator light is not illuminated

1. The circuit is not connected

2. Fused switch burnt out

Connect the circuit properly

Replace fuse

ProductsDeuterium GasPressure not available

Flow display above 300

Instrument internal airway system leak

 

Inspect all gas line connections with a detection liquid

Tighten leak point

Flow display is around 150

Please provide the Chinese content to be translated.Deuterium GasEquipment, pressure not exceeding

Instrument external air line connection leaks

ProductDeuterium gasFlow Rate

Deviation too large

Electrolytic cell aging

2. Abnormal power output current

1. Replace electrolytic cells, clean electrolytic cells

2. Adjust power supply, replace power supply

Deuterium gasImpure

Silica gel inside the drying tube has reached saturation.

Replace color-changing silicone

There is water inside the purification and drying tube.

Air-water separator cannot drain normally

Replace air-water separator

Clean separator for gas and water

Excessive consumption of pure water

1. There is a leak in the water recirculation pipeline

2. Electrolytic cell temperature too high

Observe and eliminate water leakage

Electrolytic cell aging - Replace electrolytic cell or clean electrolytic cell