I. Overview

PGD-300BDeuterium gasThe generator's electrolytic cell employs the cutting-edge solid polymer electrolyte technology for deuterium production, ensuring the outputDeuterium GasOxygen content less than 0.1ppm; it uses pure water as the electrolyte and titanium metal as the electrodes. After applying power, the cathode produces...Deuterium gasAnode produces oxygen. Oxygen is vented.Deuterium gasAfter entering the air-water separator, it separates from the water and is fully dried through the drying tube before being output from the exhaust port. This instrument employs a high-sensitivity, fuzzy automatic tracking control system, producing gas on demand and stopping production when not in use, achieving automatic constant pressure and flow. This ensures the pressure stability precision is less than 0.001 Mpa, guaranteeing the instrument's output.Deuterium GasHigh purity, pressure stable, safe, and continuous.


II. Main Technical Parameters

   Electrolytic Cell: PME Pure Water Electrolytic Cell

3、Deuterium gasFlow Rate: 0-300 ml/min

4. Output Pressure: ≥0.4 MPa

5. Available Pure Water: Water with a resistivity greater than 2MΩ, double-distilled water, or Yibao water

6. Operating Temperature: -15℃ to 50℃

7. Pressure Stability: <0.001 Mpa

8. Power Supply: 220V ±10% 50Hz

9. Power Consumption: 150W

10. Dimensions: 360mm x 190mm x 350mm

11. Sink Capacity: ≥2.0L

12. Machine Weight: Approximately 13kg

III. Causes of Failure and Troubleshooting Methods

Fault Phenomenon

Fault Cause

 

Inspection Methods | Exclusion Methods

 

Instrument cannot operate after power-on

Power indicator light is not illuminated

1. The circuit is not connected

2. Fused safety tube burned out

Connect the circuit properly

Replace Fuse

ProductDeuterium gasPressure is off.

Flow display above 300

Instrument internal air circuit system leak

 

Check all gas line connections with a testing fluid

Tighten the leak point

Flow display is around 150

Please provide the Chinese content to be translated.Deuterium gasEquipment, pressure not exceeding

Instrument external airway connection leakage

ProductDeuterium gasTraffic Value

Deviation is too great.

Electrolytic cell aging

2. Abnormal power output current

1. Replace electrolytic cells, clean electrolytic cells

2. Adjust power supply, replace power supply

Deuterium gasImpure

Silica gel inside the drying tube has reached saturation.

Replace color-changing silicone

There is water inside the purification and drying tube.

The gas-liquid separator is unable to drain normally.

Replace the air-water separator

Clean air-water separator

Excessive pure water consumption

1. Leaks in the water recirculation pipeline

2. Electrolytic cell temperature too high

Observe and eliminate leakage

Electrolytic cell aging - Replace electrolytic cell or clean electrolytic cell