详情描述

I. Composition:

LRD-SZO series Zirconia Oxygen Analyzer is a complete set of instruments that includes a zirconia oxygen sensor, oxygen transmitter, and remote display.

II. Application
LRD-SZO SeriesZirconia Oxygen Content AnalyzerRapid and accurate on-line detection and analysis of flue gas content during the combustion process of boilers, kilns, heating furnaces, and other combustion equipment to achieve low-oxygen combustion control, energy conservation, and reduce environmental pollution.
LRD-SZO SeriesZirconia Oxygen Content AnalyzerConsists of two parts: zirconium oxide heads (primary instrument) and oxygen content transmitters (secondary instrument).
The LRD-SZO type zirconia probe housing is made of high-temperature and corrosion-resistant stainless steel material. No additional gas is required.
The reference gas can self-convection. It is equipped with a standard gas interface, which allows for calibration and verification using standard gases during on-site operation. The probe zirconium tube can be easily removed and replaced.
The LRD-SZO oxygen transmitter features a simple structure, standardized installation dimensions, reasonable wiring design, advanced manufacturing quality, stable and reliable instrument performance, and easy debugging.
LRD-SZO SeriesZirconia Oxygen Content AnalyzerDue to its superior performance-to-price ratio, it has been widely used in large and medium-sized power plants domestically for several years.
Section 3: Specifications and Dimensions
1. Oxygen Transmitter Size
Platter-style Horizontal (Ⅰ) 160×80×250                152×76
Tray-mounted vertical type (Ⅱ)  - 80 × 160 × 250 or 160 - 76 × 152
Packing Method (Ⅲ) 160×160×250 or 160, 153×153
Wall-mounted (Ⅳ)   325 × 250 × 110       310 × 128
2. Zirconia probe dimensions: (in mm)

Section 4: Technical Specifications
1. Basic Tolerance: <+3% F.S.; Instrument Accuracy: Grade 1
2. Range: 0~25% O2
3. Background Correction: -20mV ~ +20mV
4. Tested flue gas temperature: ZO-4 type below 800℃ (low temperature type); ZO-5 type 800℃~1200℃ (high temperature type)
5. Output Signal: 0~10mADC, 4~20mADC, adjustable as desired
6. Load Capacity: 0~1.2KΩ (0~10mA) or 0~600Ω (4~20mA)
7. Environmental Capacity: 0~50°C, Relative Humidity < 90%
8. Power supply: 220V ±10%, 50Hz.
9. Power Consumption: Transmitter approximately 8W, heating furnace average 50W.
10. Response Time: Approximately 3 seconds, 90%.
11. Zirconia probe heating furnace heating time: approximately 20 minutes.