How to Choose an External Pump for the DR-803K Water Quality Sampler?
Publish Time:2022-04-28        View Count:53         Return to List

Cautionary Notes for Installing External Pumps

Horizontal Distance

Vertical Height

2M<Height<5Mee

5Mi < Height

20Mi > Distance >10Mi

350WSelf-Priming PumpPower Cord1SquareDC24V5ARelayControl AC Contactors, Use AC Contactors to Control PumpsThe company-provided junction box is mounted to the exterior wall of the instrument room, at a height equal to or not less than the height of the on-site distribution box.1Mi

300WSubmersible Pump; Power Cable1SquareDC24V5ARelayControl the AC contactor, use the AC contactor to control the pump.Mount to the exterior wall of the instrument room, ensuring the fixed height is level with the on-site distribution box or not less than the ground level.1Millet

Distance >20Mi

550W~750WSelf-Priming PumpPower Cord1.5SquareDC24V5ARelayControl AC Contactors, Use AC Contactors to Control Pumps

The company-provided junction boxMount the device to the outer wall of the instrument room at a fixed height that is level with the on-site electrical distribution box or not less than the ground level.1Millet

550W~750WSubmersible pump; power cable1.5SquareDC24V5ARelayControl the AC contactor, use the AC contactor to control the pump.

The company-provided junction boxMount on the exterior wall of the instrument room, at a fixed height equal to or not lower than the height of the on-site power distribution box.1Mi

Note: The power cable of the external pump must be drawn from the power distribution panel on-site, strictly forbidden to be extracted from the switch power supply of the sampler!