Logan Kexing KB-3A Portable Oil Contamination Detector

Overview:
Offline Detection:
Direct sampling for testing from the sample bottle; use a pressure bottle sampler (with pressure device) for sampling and testing.
Online Inspection:
Can be installed on various hydraulic transmission, lubrication, oil filtering machines, cleaning units, and testing stations, enabling online detection of the oil contamination (cleanliness) level of the system.
Performance Features:
Utilizes the principle of photoresist (light blocking) method, featuring fast detection speed, strong anti-interference capability, high accuracy, and good repeatability.
High-precision sensors ensure high resolution and accuracy.
High-precision sampling system, ensuring constant sampling speed and precise volume control.
4.3-inch color LCD display with touchscreen operation.
Integrated NAS1638, ISO4406-1999, ISO4406-2017, ISO4406-1987, GB/T 14039-2002, GJB420B-2006, GJB420B-2015, GJB420A-1996, AS4059E differential, AS4059E cumulative, AS4059F differential, AS4059F cumulative, ROCT17216-1971, ROCT17216-2001, JTS-005, SAE749D-1963, and NAS1638/ISO4406 particle contamination grade standards, with the ability to incorporate additional standards as per customer requirements.
In-built "Custom" standard, allows for particle size settings at intervals of 0.1μm, with up to 990 sizes available for easy particle size analysis.
In-built calibration function, capable of calibration according to standards such as ISO4402, GB/T18854-2002 (ISO11171:1999), and ROCT.
Store up to three calibration curves (ACFTD, ISOMTD, GOST) simultaneously and easily switch between them to minimize conversion errors.
Built-in printer for direct printing of test reports.
The testing data storage function allows for data to be saved either within the instrument's internal memory or to an external USB flash drive via the instrument's USB port.
The instrument can perform continuous automatic online detection, providing real-time particle count and grade, and allows for customizable detection interval settings.
Optional water activity sensor for detecting the water activity (aw value) and oil temperature in the oil.
Features an adjustable rinse function with customizable rinse volume.
High-strength injection-molded imported shell, compact structure, lightweight, allowing the main unit and accessories to be easily carried in a backpack.
High-precision aluminum anodized panel, sleek and aesthetically pleasing, durable and long-lasting.
Standard RS232 communication is provided, along with communication protocols. It can be connected to a computer or a higher-level machine to control the instrument and transmit, store, and process detection data.
Technical Specifications:
Light Source: Semiconductor Laser
Detection Range: 1μm to 600μm (depending on the selected sensor)
Particle size range: 1-100 μm (ISO 4402) or 4-70 μm (c) (GB/T 18854, ISO 11171)
l Sensitivity: 1μm (ISO 4402) or 4μm (c) (GB/T 18854, ISO 11171)
Inspection Channels: 8, particle size can be set at any interval of 0.1μm
Testing Method: Offline or Online
Sample Volume: 10ml per time
Sample Volume Relative Error: Better than ±3%
Counted Volume: 10ml
Test Time: 30S/each
Testing speed: 5-60 mL/min (default 20 mL/min)
Cleaning Speed: 5-60 mL/min (default 20 mL/min)
Flow Rate: adjustable between 0ml to 90ml (in increments of 1ml)
Overlap error limit: 12,000 to 40,000 particles/mL (depending on the selected sensor)
Particle count repeatability: RSD < 2%
Particle Count Relative Error: ±10%
Results Storage: Inside the instrument or on a USB drive
Offline viscosity test: ≤100cSt (optional pressure bottle sampler can reach 400cSt)
Online Pressure Test: 0.1~0.6 Mpa (optional pressure reducer for pressure < 40 MPa)
Online Testing Interval: Can be set between 1 second and 10 hours, 59 minutes, and 59 seconds.
Test Sample Temperature: 0℃~80℃
Operating Temperature: -20℃ to 60℃
Storage Temperature: -30℃ to 80℃
Power supply: 100V~265VAC or 24VDC or external lithium battery pack
Dimensions: 345×295×152mm
Instrument Net Weight: 5 kg