Logan Kexing KT-2A Desktop Liquid Particle Size Analyzer | Particle Counter

Developed based on the principle of photoresist (light blocking) method, used for measuring the size and quantity of solid particles in liquids.,Widely applicable in the aviation industry、AerospaceAviation engines, heavy industry machinery.In the fields of power, oil, petrochemicals, transportation, ports, metallurgy, machinery, and automotive manufacturing, solid particle contamination testing is conducted on fluids such as hydraulic oil, lubricating oil, rock oil, transformer oil (insulating oil), turbine oil, gear oil, engine oil, aviation kerosene, and water-based hydraulic oil.And regardingWater-based samples, chemical reagents,Organic liquids and polymer solutions for insoluble particle treatmentSampleInspection.
Performance Features:
l Utilizing the principle of photoresist (light shielding) method, it boasts advantages such as fast detection speed, strong anti-interference capability, high accuracy, and good repeatability.
l High-precision sensors ensure high resolution and accuracy.
l Precision plug-type sampling system, ensuring consistent sampling speed and precise volumeSureControl.
l Please provide the Chinese content to be translated into American English./Vacuum pressure vessel device for degassing samples and testing high viscosity samples.
l 5.6InchColor LCDScreen DisplayTouchScreenOperation.
l Built-inNAS1638、ISO4406-1999、ISO4406-2007、ISO4406-1987、GB/T 14039-2002、GJB420B-2006、GJB420B-2015、GJB420A-1996、AS4059E、AS4059F、ROCT17216-1971、ROCT17216-2001、JTS-005、SAE749D-1963、NAS1638/ISO4406Particle contamination grade standards available, and can be fitted with required standards upon customer request.
l Built-in"Customized" standards, with optional spacing0.01μmCan be set to any particle size dimension.9900Particle size, for easy particle size analysis.
l Can store up to four calibration curves simultaneously (latex ball calibration curve,ACFTDCalibration CurveISOMTDCalibration curvesGOSTCalibration curve), and can be easily switched, reducing conversion errors.
l Test data storage function for easy archiving, retrieval, and analysis of test data.
l Built-in printer, capable of directly printing out test reports.
l Built-in Chinese input method, enabling Chinese annotations on inspection reports.
l RS232Interface and powerful data functionalitiesHarvestSoftware for external computers to process detection dataLabelingStorage,SearchGeneratePDFInspection reports, etc.
l Features magnetic stirring for even particle distribution.
l The pressure pump is equipped with four-stage gas filtration and a water purification unit, preventing environmental air from contaminating the samples.

Technical Specifications:
1.Light Semiconductor Laser
2.Scope of Inspection:0.8μm~600μm(Subject to the selected sensor)
3.Particle size range:1~100μm( ISO4402) or4~70μm(c) (GB/T18854,ISO11171)
4.Spirit Sensitivity Degree:0.8μm(ISO4402)or3μm(c) (GB/T18854,ISO11171)
5.Inspection Channel:16Pieces, with intervals0.01μmAny particle size can be set
6.Sample Volume:0.2~100mlComma spacing0.1ml
7.Relative Error of Sample Volume: Better than ±0.5%
8.Testing Speed:5~80mL/min(Default)15mL/min)
9.Cleaning Speed:5~80mL/minDefault15mL/min)
10.Please provide the Chinese content to be translated. Identify Power: Superior10%(GBT18854-2002,ISO11171)
11.Overlap Error Limit:12000~40000Grain/mL(Subject to the selected sensor)
12.Particle count repeatability:RSD<2%
13.Particle Count Relative Error: ±5%
14.Pressure Vessel Negative Pressure<-0.08MPa
15.Positive pressure in the pressure chamber:<1MPa
16.Test Sample Viscosity: ≤650cSt
17.Inspect Sample Temperature:0℃~80℃
18.Operating Temperature:0℃~60℃
19.Storage Temperature:-30℃~80℃
20.Supply Electricity100~245V,49~62Hz
21.Dimensions: Main Machine 430×320×590mm Air Pump 330×266×255mm
21.Instrument Net Weight: Main Unit30kgPressure Pump10kg