Part 1: System Composition Overview
The system is composed of the following components:
The function of the test chamber is to provide a stable environment with constant high and low temperature and humidity, as well as alternating high and low temperature and humidity; it includes a refrigeration system, heating system, humidification/dehumidification/waterway system, air duct system, control system, and system safety protection devices.
Part 2:Main Technical Specifications of the Test Box
1. Studio size: 1000×1000×1000mm (width x depth x height)
2. Dimensions approximately: 1510×1560×1860mm (width×depth×height)
3. Temperature range: -40°C to +150°C
4. Temperature fluctuation: ≤±0.5℃
5. Temperature uniformity: ≤±1℃
6. Temperature Deviation: ≤±2℃
7. Rate of temperature rise/fall
From +20°C to -40°C, approximately 60 minutes
From +20 to 100°C, approximately 40 minutes
8. Humidity Range: 20% - 98% R.H. (AT +25℃ to +85℃)
9. Humidity Error: +2/-3% R.H. (above 75% R.H.), ±5% R.H. (below 75% R.H.)
10. Wind Speed: 1.7 ~ 2.5 m/s
These metrics were obtained under conditions of an ambient temperature ≤25°C, atmospheric pressure, and with the container unloaded and unburdened, within a 1/6th space from the inner wall of the container.
11. Power: approximately 12.5 kW
12. Power: 380V ± 10%V; 50Hz
Comply with relevant standards and testing methods:
GB/T2423.22-87Nb GJB1032-90 MIL-STD-2164/(E/C)
GB/T2423.4-93 GB/T2423.34-86 GJB150.9-86