CCZ20Mineral dust sampler
The CCZ20 Dust Collector is an instrument used for measuring the concentration of suspended dust in ambient air. The instrument conforms to the GB/T 20964-2007 "General Technical Conditions for Dust Samplers" and the JJG520-2005 "Routine Inspection Procedures for Dust Samplers."Q/320582YLA00Model 1-2010Dust Sampler Enterprise StandardExibI explosion-proof design in accordance with GB3836.1-2000 and GB3836.4-2000 standards, suitable for coal mine underground and other workplaces with explosive gas hazards.
This sampler is equipped with two dust precollection filters: one for measuring the total dust concentration throughout the process, and the other for measuring the respirable dust concentration. This precollection filter can separate harmful respirable and non-respirable dust particles, allowing for the collection of both respirable and non-respirable dust samples in a single collection, with a separation efficiency that meets internationally recognized standards."BMRC" curve standard is a relatively reliable dust separation device. The instrument is equipped with a high-performance suction pump featuring pulseless airflow, with large load and vacuum capacity, automatic timing sampling, stable flow rate, easy operation, and portability.
Key Technical Specifications:

1 Sample Flow Rate: Total Dust Sampling5-30 L/min, 5-25 L/min, 10-25 L/min, 1-10 L/min (optional)
During respirable dust sampling:20 L/min (Flow Rate Fixed))
2 Sample Flow Error:≤3%FS
3 Sampling Flow Stability: SamplerSample flow stability ≤ 3.0% FS within 60 minutes
4 Load Capacity:≥200Pa
5 Continuous Operation Time:≥100min
6 Work noise:≤75dB (A)
7 Flow meter accuracy: Not less than2.5 grade.
8 Sampling Time Error≤±0.1% within 5 minutes
9 Average Sampling Flow Rate:0.5~4.0m/s
10 Accuracy of sampling:≤10%
11 Airtightness: The sampler's sampling head and flowmeter maintain a pressure difference of (1000-1100) Pa for 1 minute, with pressure change not exceeding 100 Pa.
Operating Environment Temperature: -10~40℃ Relative Humidity: ≤95%
13 Atmospheric Pressure: 80KPa to 110KPa - Coal mine with potential for gas or coal dust explosion hazards underground
14 Dust Collection Range: All dust, respirable dust.
15 Explosion-proof Type: Mining intrinsic safety type ExibI (150℃), suitable for flammable and explosive workplaces and underground coal mine sampling.
16 AA Battery Pack: Ni-H Nickel Hydride Battery Cells7A/h, 6 sections, in series with 4 0.51Ω/10W current-limiting resistors, encapsulated in epoxy and placed inside the battery box.Rated VoltageUe:7.2V;High output voltageU0:9.0V; Output Current I0:4.52A。
17 Dimensions: 149×202×80 (mm)
18 Weight ≤ 1.75 Kg
19 Sampling Efficiency: Meets "B" standardMRCSample Efficiency 4, the sample efficiency allows for errors in accordance with the specifications in Table 1.
Table1 Respirable dust sampler sampling efficiency allows for error
Particle size/μm |
| 5.9 | 5.0 | 3.9 | 2.2 |
Sample Efficiency% | 0 | 30 | 50 | 70 | 90 |
Sampling efficiency allows for error% |
|
|
|
|
|
T